Under-clearing

Also known as: incomplete clearing, cloudy on infiltration

What it looks like

Poor wax infiltration and soft blocks because alcohol was not fully replaced by clearant before wax.

Where it usually appears: Short clearing steps, exhausted clearant, or fatty tissue.

Likely causes
  • Insufficient clearing time
  • Exhausted or diluted clearing agent
  • Residual water/alcohol from incomplete dehydration
  • Fatty tissue needing more clearing
Immediate checks
  • Check clearant rotation and change overdue stations
  • Confirm dehydration was complete
  • Extend clearing for fatty tissue
Corrective action
  • Reprocess affected tissue through fresh clearant and wax
Preventive action
  • Maintain clearant change schedule
  • Ensure full dehydration first
  • Longer schedule for fatty/dense tissue
Equipment: Tissue processorReagent: Clearing agentRelated SOP: Tissue processing SOP

This atlas is for educational purposes for laboratory professionals. It does not replace your institutional SOPs, manufacturer instructions, validated protocols or pathologist judgment. Adapt any action to your laboratory and equipment.

Unimeditrek supports histopathology laboratories with tissue processors, microtomes, staining and workflow equipment plus service and AMC. Talk to a specialist →

Request a bulk quote

Tell us the consumables & quantities you need — we'll send a GST proforma. No payment now.

×Unimeditrek histopathology equipment — enlarged view
Click image to zoom · click outside to close
Verified download

Download document

Verify your details and we'll send a one-time code to your email and mobile. Your details are added to our enquiry desk so our team can help if you need it.

WhatsApp +91-9810533445 · 24/7 Request a Quote