Cryostat section curling

Also known as: sections roll up, won't flatten on the anti-roll plate

What it looks like

Sections curl or roll instead of lying flat as they are cut in the cryostat.

Where it usually appears: Anti-roll plate misaligned, wrong temperature, or static.

Likely causes
  • Anti-roll plate misaligned or dirty
  • Cryostat/tissue temperature not optimal for the tissue
  • Blade angle incorrect
  • Static electricity
Immediate checks
  • Adjust and clean the anti-roll plate to sit just above the blade edge
  • Adjust cabinet temperature for the tissue (fatty tissue colder)
  • Check blade angle
Corrective action
  • Realign the anti-roll plate and re-cut
  • Use a brush to guide flat sections if needed
Preventive action
  • Routine anti-roll plate care
  • Set temperature by tissue type
  • Reduce static
Equipment: Cryostat microtomeRelated SOP: Cryostat frozen section SOP

This atlas is for educational purposes for laboratory professionals. It does not replace your institutional SOPs, manufacturer instructions, validated protocols or pathologist judgment. Adapt any action to your laboratory and equipment.

Unimeditrek supports histopathology laboratories with tissue processors, microtomes, staining and workflow equipment plus service and AMC. Talk to a specialist →

Request a bulk quote

Tell us the consumables & quantities you need — we'll send a GST proforma. No payment now.

×Unimeditrek histopathology equipment — enlarged view
Click image to zoom · click outside to close
Verified download

Download document

Verify your details and we'll send a one-time code to your email and mobile. Your details are added to our enquiry desk so our team can help if you need it.

WhatsApp +91-9810533445 · 24/7 Request a Quote