UNIMEDITREKTOTAL MEDICAL SOLUTIONS
SOP TEMPLATE
Doc ref: UMT/SOP/H-AND-E-STAINING
Reviewed: July 2026
Rev.: template

H&E Staining SOP

Histopathology · Unimeditrek Pvt. Ltd.
⚠️ Validation required. This SOP template is provided for educational and quality-system support only. It is a starting point to adapt — not a validated procedure. Each laboratory must review, complete the lab-specific fields, validate and approve this SOP through its own pathologist / quality manager, and align it with local regulatory requirements before use.

1. Purpose

To produce reproducible hematoxylin and eosin stained sections with crisp nuclear detail and clear cytoplasmic contrast, suitable for routine histopathology reporting.

2. Scope

Applies to routine H&E staining of paraffin sections, manual or automatic. Special stains are covered by their own SOPs.

3. Responsibility

4. Equipment & materials

5. Safety precautions

6. Procedure

Deparaffinise & rehydrate

  1. Remove paraffin in xylene/substitute, then rehydrate through descending graded alcohols to water.
  2. Ensure deparaffinisation is complete — residual wax blocks aqueous staining. Bath times: [lab to validate].

Nuclear stain

  1. Stain nuclei in hematoxylin. Time: [lab to validate] (depends on hematoxylin type and freshness).
  2. Rinse, then differentiate briefly in acid alcohol to sharpen nuclear detail — avoid over-differentiation.
  3. Blue the sections adequately (bluing reagent or running water). Time: [lab to validate].

Counterstain & finish

  1. Counterstain cytoplasm in eosin. Time: [lab to validate]; avoid over-rinsing which strips eosin.
  2. Dehydrate through ascending alcohols, clear in xylene/substitute and coverslip promptly (do not let sections dry before mounting).

7. Quality-control checkpoints

8. Acceptance criteria

9. Common errors & corrective action

ProblemCorrective action
Weak / pale nucleiReplace exhausted hematoxylin; reduce differentiation; ensure adequate bluing and complete deparaffinisation; cut thinner sections.
Over-dark nucleiReduce hematoxylin time and/or increase differentiation slightly; standardise on programmed times.
Uneven / patchy stainingStandardise agitation and reagent levels; an automatic stainer removes manual variation.
Precipitate on slideFilter/skim oxidised hematoxylin and clean staining containers.
Tissue lifting during stainingUse adhesive/charged slides and ensure adequate slide drying; moderate agitation.

10. Records & documentation

11. Review frequency

Review at least annually, or after any reagent, protocol or equipment change, or a recurring defect.

12. Approval

RoleNameSignatureDate
Prepared by
Reviewed by (Quality Manager)
Approved by (Pathologist)
Unimeditrek Pvt. Ltd. — histopathology laboratory equipment manufacturer, India. Author: Unimeditrek Technical Content Team · Reviewed by: Biomedical Engineering & Histopathology Applications team · Last reviewed: July 2026.
📞 +91-98105 33445 · sales@unimeditrek.com · www.unilabequipments.com · Manufactured under ISO 9001 & ISO 13485 QMS; certificates on request.
© Unimeditrek Pvt. Ltd. This is an adaptable template for laboratory use — complete the highlighted fields and validate through your own quality system before implementation.